Model Number N/A |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that debris was identified in the sterile packaging while investigating items in stock.No patients were involved.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d9; g3; h2; h3; h6 complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned product found an unknown piece of debris within the sterile packaging.The debris was sent for additional analysis.The ftir analysis identified the debris as polypropylene.Review of the device history record(s) identified no deviations or anomalies during manufacturing.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|