ZIMMER BIOMET, INC. G7 STR INSRTR THREADED SHAFT; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a primary case there was an issue with the g7 straight handle inserter and the threads came off during impaction.To complete the procedure a curved impactor was used.There was no delay in result of the broken straight handle.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No additional information on the reported event.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed via pictures.Visual examination of the provided photos identified an inserter with broken threaded end displayed as well as attempt to show item and lot number that was blurry.No product was returned; dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR INSRTR THREADED SHAFT
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11076611
MDR Text Key223767994
Report Number0001825034-2020-04424
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003452
Device Lot Number415830
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberNI
Patient Sequence Number1
-
-