HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0; VENTRICULAR (ASSIST) BYPASS
|
Back to Search Results |
|
Model Number 1420 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the controller had a loose driveline port connector.The controller remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for additional information and correction.Manufacturer name was corrected from (b)(4) to heartware, inc.Manufacturer fax no.Was corrected from blank to (b)(4).Manufacturer email was corrected from blank to (b)(4).Unique identifier (udi) # was corrected from asku to (b)(4).Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for device evaluation.Product event summary: the controller was not returned for evaluation.A review of the manufacturing documentation confirmed that the associated device met all requirements for release.Visual evidence provided by the site revealed that the controller¿s driveline connector was loose.As a result, the reported event was confirmed.Capa00230 investigated loose driveline connectors and determined the root cause to be a design issue, where the thread locker did not have sufficient strength to secure the nut.Corrective actions implemented included the incorporation of an improved thread locker.The most likely root cause is inadequate thread locking resulting from the low bond strength of the adhesive.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|