Model Number 71935104
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during surgery the trial broke upon removal after trailing components.No delay or injury reported.It is unknown how the procedure was finished.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Event Description
It was reported that during surgery the trial broke upon removal after trialing components.All pieces were recovered.No delay reported.The procedure was completed as normal with the intended prothesis being implanted.Trialing was complete at the time of breakage.There was no need for a backup device.No changes to technique were needed.
 
Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is broken into several pieces, rendering the device inoperable.The clinical/medical investigation concluded that per complaint details, the trialing was complete at the time of breakage and the procedure was completed as planned without the need for a backup device or change in surgical technique.No injury or delay was reported, therefore, no further medical investigation is warranted.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JII UNI INS TRIAL SZ 3-4 8MM
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11089416
MDR Text Key224133576
Report Number1020279-2020-07925
Device Sequence Number1
Product Code HSX
UDI-Device Identifier00885556577677
UDI-Public00885556577677
Combination Product (y/n)N
PMA/PMN Number
K152315
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number71935104
Device Catalogue Number71935104
Device Lot Number16KAP9750
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2020
Date Manufacturer Received07/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-