ST. JUDE MEDICAL FLEXABILITY ABLATION CATHETER, SENSOR ENABLED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number A-FASE-FJ |
Device Problem
Signal Artifact/Noise (1036)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/11/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Related manufacturer report number: 3005334138-2020-00654.During a vt ablation procedure, noise was noted on the catheter.A second device was attempted, but the same issue occurred.A third device was attempted, and the noise was still present.However, the procedure was continued with the third catheter with the noise present and no adverse consequences to the patient.A clinically significant delay occurred and the patient was given additional anesthesia during the procedure due to this issue.
|
|
Manufacturer Narrative
|
Additional information: d9, g3, g6, h2, h3, h6.One bi-directional, curve f-j, tacticath sensor enabled contact force ablation catheter was received for evaluation.Electrical testing of the returned device determined electrodes 1-4 met specifications for acceptable resistance values with no open or short circuits detected.A simulated ablation test was conducted; however, noise was not present on the distal electrode before, during, or after ablation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported noise issue remains unknown.
|
|
Search Alerts/Recalls
|
|
|