• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Power Problem (3010)
Patient Problem No Patient Involvement (2645)
Event Date 08/12/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Telephone number: (b)(6).Product evaluation investigation is complete.This is an initial final report submission.Review of the most recent repair record determined the unit was out of calibration, the machined head was damaged and the motor had low rpm.The unit was recalibrated, and the machined head and motor were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported that during surgery, the device does not function at all.Alternate device used without consequences to the patient or the surgeon.There was a delay of 0-15 minutes.Product evaluation investigation determined the device had low rpm; a reportable malfunction.No adverse events were reported as a result of this malfunction.No additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11091691
MDR Text Key224398529
Report Number0001526350-2020-01123
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number61632114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/18/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-