V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Defective Alarm (1014); Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 29dec2020.
 
Event Description
It was reported to philips light emitting diode (led) alarm failed inoperable (inop) being displayed.Diagnostic code consistent with led alarm failure was present.The remote service engineer (rse) provided the biomed with service exchange part number for power switch overlay power management board, switch printed circuit board assembly, cable to power management and price so they can order direct and install themselves and biomed also provided with fixed price repair details.The device did not have patient involvement at the time the issue was discovered, therefore, there was no patient or user harm reported.
 
Manufacturer Narrative
G4:15mar2021.B4:16mar2021.Diagnostic code consistent with led alarm failure was present.The remote service engineer (rse) provided the biomed with the service exchange part number for power switch overlay, power management board, switch printed circuit board assembly, cable to power management, and price to order directly and install themselves.The biomed was also provided the fixed price repair details.The customer was provided with the parts required, but as the device is no longer under contract, it is up to the customer if/when they carry out the repair.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11092149
MDR Text Key225304521
Report Number2031642-2020-04792
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/26/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-