• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO PREMIUM SURGICLIP III; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO PREMIUM SURGICLIP III; CLIP, IMPLANTABLE Back to Search Results
Model Number 133650
Device Problems Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during an open abdominal aortic replacement, on blood vessel ligature, the device was able to be used without any problem immediately after use; however, a clip could not be accurately formed during clipping.A malformed clip fell into the patient's cavity, but it was visually checked and removed completely.After that, a new product was taken out and the procedure continued.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the clips did not close completely and were malformed.A component disengaged from the device into the surgical cavity.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: do not excessively twist or torque the jaws.Excessive twisting or torqueing the jaws may dislodge clip from the jaws or cause clip malformation after jaw closure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREMIUM SURGICLIP III
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key11094525
MDR Text Key224280062
Report Number2647580-2020-04247
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521517394
UDI-Public10884521517394
Combination Product (y/n)N
PMA/PMN Number
K142869
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number133650
Device Catalogue Number133650
Device Lot NumberP0D1556Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2020
Date Manufacturer Received02/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-