EDWARDS LIFESCIENCES INSPIRIS RESILIA AORTIC VALVE; TISSUE, HEART-VALVE Back to Search Results
Model Number 11500A25
Device Problem Insufficient Information (3190)
Patient Problem Insufficient Information (4580)
Event Date 12/01/2020
Event Type  Injury  
Manufacturer Narrative
Udi number: (b)(4).Bioprosthetic tissue valves can deteriorate with time and eventually fail contributing to regurgitation and/or stenosis.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular regurgitation and/or stenosis.In this case, minimal information regarding this valve replacement procedure was received and attempts to obtain additional information regarding the condition of the device, patient's medical history, or possible comorbidities have been made.The root cause of this event cannot be determined with the available information.However, this event was most likely impacted by the progression of the patient¿s underlying valvular disease pathology with or without structural valve deterioration and/or nonstructural dysfunction.Requests for additional information have been performed.The healthcare provider has neither returned the explanted valve nor provided any additional information at this time.The device was not returned for evaluation.If returned, a supplemental report including the evaluation findings will be submitted.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.If new information becomes available, a supplemental report will be submitted.
 
Event Description
It was learned via the patient registry that a patient with a 25mm 11500a pericardial aortic valve underwent a redo aortic valve replacement procedure after an implant duration of four (4) months due to unknown reasons.The explanted valve was replaced with a 23mm 11500a pericardial valve.
 
Manufacturer Narrative
Updated: b4, g4, h6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRIS RESILIA AORTIC VALVE
Type of Device
TISSUE, HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key11096523
MDR Text Key224798775
Report Number2015691-2020-15331
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
PMA/PMN Number
P150048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number11500A25
Was Device Available for Evaluation? No
Date Manufacturer Received01/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
-
-