Model Number N/A |
Device Problem
Intermittent Loss of Power (4016)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/08/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Telephone number (b)(6).Once the investigation is complete, a follow up/final report will be submitted.
|
|
Event Description
|
It was reported that on the day of the procedure the dermatome works intermittently with the power going on and off.An alternate device was used for the procedure.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the rpms were out of specification on the low end.The motor was replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|
|
Event Description
|
No additional event information.
|
|