• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem Low Readings (2460)
Patient Problem Hyperglycemia (1905)
Event Date 12/22/2020
Event Type  Injury  
Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device manufacturing date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A caller reported that the customer had a low readings issue with the freestyle libre 2 sensor.The caller reported unspecified low readings, and the customer self-treated with food.As the customer did not have test strips, she had contact with a healthcare professional.A healthcare meter reading of 150 mg/dl was obtained, as compared to a sensor scan of 80 mg/dl.The healthcare professional replaced the sensor and gave the customer an insulin injection (dose/type unknown).There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Sensor (b)(6) has been returned and investigated.The sensor plug is fully seated and no issues were observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Inspected the plug assembly, no issues were observed.The current was applied to the sensor to perform linearity testing while in the test fixture.All results were within specification.The issue is therefore not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A caller reported that the customer had a low readings issue with the freestyle libre 2 sensor.The caller reported unspecified low readings, and the customer self-treated with food.As the customer did not have test strips, she had contact with a healthcare professional.A healthcare meter reading of 150 mg/dl was obtained, as compared to a sensor scan of 80 mg/dl.The healthcare professional replaced the sensor and gave the customer an insulin injection (dose/type unknown).There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
No product has been returned.Extended investigation has been performed for the reported complaint.And there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed, and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A caller reported, that the customer had a low readings issue with the freestyle libre 2 sensor.The caller reported, unspecified low readings.And the customer self-treated with food.As the customer did not have test strips.She had contact with a healthcare professional.A healthcare meter reading of 150 mg/dl was obtained, as compared to a sensor scan of 80 mg/dl.The healthcare professional replaced the sensor.And gave the customer an insulin injection (dose/type unknown).There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key11097714
MDR Text Key224511606
Report Number2954323-2020-14655
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup
Report Date 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/31/2022
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2021
Date Manufacturer Received01/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient Weight95
-
-