Model Number 1458Q/86 |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/28/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
This product is registered as a combination product.Further information was requested but not received.The results, method, and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During an in-clinic follow up, lead dislodgement was noted on the left ventricular (lv) lead.The lv lead was explanted and replaced to resolve the event.The patient was stable with no consequences.
|
|
Manufacturer Narrative
|
The reported event was lead dislodgement.As received, a complete lead was returned in one piece.The s-curve hump height was measured within specification.Visual examination did not find any anomalies.Electrical testing did not find any indication of conductor fractures or internal shorts.
|
|