INTEGRA LIFESCIENCES(IRELAND) MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
During a scheduled in-service by an integra sales representative, it was observed that the index knob of the mayfield skull clamp was not functioning properly, was difficult to adjust and moved from the locked to unlocked position.There was no patient involvement, thus no delay in surgery.
 
Manufacturer Narrative
Udi: (b)(4).Device history record - the dhr shows no abnormalities related to the reported failure.The reported complaint was confirmed from the evaluation of the returned a1059 mayfield skull clamp.Evaluation showed that lock has rotational and lateral movement and a residue buildup is present.The unit needs repair, pm and replacement of worn parts.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES(IRELAND)
ida business&technology park
ida business&technology park
sragh, tullamore, co.offaly
EI 
MDR Report Key11100347
MDR Text Key224535790
Report Number3004608878-2020-00769
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2020
Date Manufacturer Received01/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-