Catalog Number A1059 |
Device Problems
Material Too Rigid or Stiff (1544); Naturally Worn (2988)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported that the rocker arm of the mayfield skull clamp (a1059) was stiff, the ratcheting mechanism was not smooth and starbursts are worn down.It is unknown if there was patient involvement or if the device failure led to patient injury or surgical delay.
|
|
Manufacturer Narrative
|
Udi: (b)(4).Mayfield skull clamp (a1059) was returned for evaluation.Evaluation shows that unit needs preventive maintenance, repair and replacement of worn parts.The reported complaint was confirmed.The device exceeds its expected life of 7 years (manufactured in 2010).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|