INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problems Material Too Rigid or Stiff (1544); Naturally Worn (2988)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the rocker arm of the mayfield skull clamp (a1059) was stiff, the ratcheting mechanism was not smooth and starbursts are worn down.It is unknown if there was patient involvement or if the device failure led to patient injury or surgical delay.
 
Manufacturer Narrative
Udi: (b)(4).Mayfield skull clamp (a1059) was returned for evaluation.Evaluation shows that unit needs preventive maintenance, repair and replacement of worn parts.The reported complaint was confirmed.The device exceeds its expected life of 7 years (manufactured in 2010).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key11102124
MDR Text Key226095653
Report Number3004608878-2020-00784
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Device Lot Number101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2020
Date Manufacturer Received01/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-