Model Number 0675 |
Device Problems
Signal Artifact/Noise (1036); Pacing Problem (1439); Pocket Stimulation (1463); Low impedance (2285)
|
Patient Problems
Undesired Nerve Stimulation (1980); No Code Available (3191)
|
Event Date 09/20/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).The lead has been returned for analysis.This report will be updated upon completion of analysis.
|
|
Event Description
|
It was reported that this patient was in the following physicians clinic and this right ventricular (rv) lead was exhibiting a low out of range shock impedance measurement of zero (0), a pace impedance measurement of 259 ohms and the device pocket was exhibiting pocket stimulation.The patient is device therapy dependent.Boston scientific technical services (ts) indicated that there was a previous stored episode in which the associated implantable cardioverter defibrillator (icd) device provided 5 shocks for a true ventricular tachycardia (vt) and the shock impedance was 39 ohms for the delivered shocks.In addition, there was another stored episode that exhibited noise with a 5 second pause in pacing therapy.The current rhythm was noted to be a normal sinus rhythm at a rate of 70 beats per minute (bpm), but the device is still pacing at the lower rate limit of 50 bpm.Ts indicating that it appears the lead is no longer sensing appropriately.Ts discussed with the boston scientific representative (rep) that they are concerned that there is a lead insulation breach, the patient should be considered unprotected and the lead needs to be replaced.The rv lead was subsequently explanted and replaced.No additional adverse patient effects were reported.
|
|
Event Description
|
It was reported that this patient was in the following physicians clinic and this right ventricular (rv) lead was exhibiting a low out of range shock impedance measurement of zero (0), a pace impedance measurement of 259 ohms and the device pocket was exhibiting pocket stimulation.The patient is device therapy dependent.Boston scientific technical services (ts) indicated that there was a previous stored episode in which the associated implantable cardioverter defibrillator (icd) device provided five (5) shocks for a true ventricular tachycardia (vt) and the shock impedance was 39 ohms for the delivered shocks.In addition, there was another stored episode that exhibited noise with a five (5) second pause in pacing therapy.The current rhythm was noted to be a normal sinus rhythm at a rate of 70 beats per minute (bpm), but the device is still pacing at the lower rate limit of 50 bpm.Ts indicating that it appears the lead is no longer sensing appropriately.Ts discussed with the boston scientific representative (rep) that they are concerned that there is a lead insulation breach, the patient should be considered unprotected and the lead needs to be replaced.The rv lead was subsequently explanted and replaced.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Patient code 3191 captures the reportable event of surgery.The lead has been returned for analysis.This report will be updated upon completion of analysis.Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner/outer insulation integrity.Measurements were within normal limits, except the pressure test which failed due to a cut on the outer insulation which likely occurred during the explant procedure.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no additional anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.
|
|
Search Alerts/Recalls
|