• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PRIMUS INFINITY EMPOWERED; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PRIMUS INFINITY EMPOWERED; ANESTHESIA UNITS Back to Search Results
Catalog Number 8607500
Device Problems Gas Output Problem (1266); Failure to Deliver (2338); Protective Measures Problem (3015)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/14/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event Description
It was reported that there was a ventilator failure during use.There was no patient injury reported.
 
Manufacturer Narrative
The ventilator failure could be reconstructed by means of the logfile analysis.No indications for a device malfunction were found.The autonomous safety shutdown of the ventilator was triggered by a significantly increased airway pressure peak.Typical reason for this is an existing lack of fresh gas in combination with spontaneous breathing efforts of the patient.However, in the course of investigation, the user denied that the patient did breathe spontaneously or that suction was used.Nevertheless, no indications for a technical device failure were found ¿ the replaced motor unit was tested in a lab device and no deviations from specification were found.A test run for several days was successfully completed without any findings.Thus, the exact root cause for the found instable pressure situation cannot be determined finally.It can be concluded that the device in question behaved as specified upon the detected pressure situation with an autonomous shutdown while changing mode to man/spont (safety mode) accompanied by an audible and visible "ventilator fail" alarm.The number of similar cases, related to the same phenomenon, is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
It was reported that there was a ventilator failure during use.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMUS INFINITY EMPOWERED
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM  23542
MDR Report Key11116017
MDR Text Key228946942
Report Number9611500-2021-00001
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8607500
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-