Product analysis for the lead was completed and approved.An analysis was performed on the returned lead portion.During the visual analysis the (+) white and (-) green helices appeared tangled; the green (-) electrode ribbon has a crease.This most likely occurred due to manipulation of the lead during the attempted implant procedure.The anchor tether, (+) white and (-) green electrodes were mounted onto lead training fixture and no issues were observed.The abraded insulation allegation was not confirmed.However, a cut on (-) tubing and coil was observed.The cut appears to have been made by a sharp object which could have occurred during the attempted implant/explant procedure; however, this cannot be confirmed.The condition of the returned lead assembly is consistent with conditions that typically exist following an attempted implant procedure.Other than the noted tangled helices, and (-) tubing and coil cut, no other obvious anomalies were noted.
|