SMITH & NEPHEW, INC. REF INTERFIT THRD HOLE COVER; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
|
Back to Search Results |
|
Model Number 71330001 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference case (b)(4).
|
|
Event Description
|
It was reported that, during an unknown procedure, the cap does not screw in properly.Instruments inside the patient.Procedure was finished with a back up device.
|
|
Manufacturer Narrative
|
H6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device has signs of damage from the attempted insertion.The dimensional inspection found the device to be within specification.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|