MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 135 DEG DHS PLATE-SHORT BARREL 2 HOLES/46MM-STERILE; DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 135 DEG DHS PLATE-SHORT BARREL 2 HOLES/46MM-STERILE; DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT Back to Search Results
Model Number 281.520S
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  Injury  
Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Manufacturing location: (b)(4).Released to warehouse: 06-feb-2019.Expiration date: 31-dec-2027.Part number: 281.520s, 135 deg dhs plate-short barrel 2 holes/46mm-sterile.Lot number: h807562 (sterile).Lot quantity 24.This dhr review is for monument operations only: thermal rinse, packaging / labeling, sterilization and release work order traveler met all inspection acceptance criteria.Packaging label log (pll) was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.(b)(4) was reviewed and determined to be conforming.This lot met all visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.This lot met all visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2020 the patient underwent a hardware removal of the dynamic hip system (dhs) from the left femur due to the patient's age.The plate, lag screw, and (2) 4.5 cortex screws were removed without issues.The procedure was successfully completed.There was no surgical delay reported.Patient outcome was unknown.This report is for one (1) 135 deg dhs plate-short barrel 2 holes/46mm-sterile.This is report 2 of 4 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
135 DEG DHS PLATE-SHORT BARREL 2 HOLES/46MM-STERILE
Type of Device
DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key11130742
MDR Text Key225496043
Report Number2939274-2021-00035
Device Sequence Number1
Product Code JDO
UDI-Device Identifier10886982177551
UDI-Public10886982177551
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K791619
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number281.520S
Device Catalogue Number281.520S
Device Lot NumberH807562
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/06/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-