• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 806455
Device Problem Mechanical Problem (1384)
Patient Problems No Known Impact Or Consequence To Patient (2692); Insufficient Information (4580)
Event Date 12/16/2020
Event Type  malfunction  
Manufacturer Narrative
Per the facility's distributor, the end user opened the lid and used a clamp.Several attempts were made to use the occluder to close the line but it would not work.
 
Event Description
It was reported that the venous occluder would not close 100%.There was no delay.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) observed the occluder plunger to not consistently completely occlude the tubing.The pst performed verification/release testing 15 times and the occluder head failed to occlude 100% 10 times, ranging from a slow drip to a fine stream.
 
Manufacturer Narrative
The reported complaint was confirmed.The service repair technician determined that the user had already been provided a replacement occluder so no further testing and/or repairs were completed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key11132314
MDR Text Key225554456
Report Number1828100-2021-00008
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup,Followup
Report Date 05/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number806455
Device Catalogue Number806455
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2021
Was the Report Sent to FDA? No
Date Manufacturer Received05/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-