Model Number 806455 |
Device Problem
Mechanical Problem (1384)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); Insufficient Information (4580)
|
Event Date 12/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the facility's distributor, the end user opened the lid and used a clamp.Several attempts were made to use the occluder to close the line but it would not work.
|
|
Event Description
|
It was reported that the venous occluder would not close 100%.There was no delay.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) observed the occluder plunger to not consistently completely occlude the tubing.The pst performed verification/release testing 15 times and the occluder head failed to occlude 100% 10 times, ranging from a slow drip to a fine stream.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The service repair technician determined that the user had already been provided a replacement occluder so no further testing and/or repairs were completed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|