Model Number 801041 |
Device Problems
Mechanical Problem (1384); Difficult or Delayed Activation (2577)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) could not verify the reported issue.The fsr tested and ran the roller pump per the preventative maintenance (pm) procedure.The suspect device was returned to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that the roller pump had hesitation when starting the revolutions per minute (rpms) and also had a grinding noise.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician observed the roller pump to operate as intended throughout the evaluation.There was not any noted hesitation when starting or running the pump.
|
|
Manufacturer Narrative
|
The reported complaint could not be confirmed.During testing of the device at the service center, the service repair technician observed the roller pump to function properly.Preventive maintenance and release testing was performed.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|