Model Number 801041
Device Problems Mechanical Problem (1384); Difficult or Delayed Activation (2577)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) could not verify the reported issue.The fsr tested and ran the roller pump per the preventative maintenance (pm) procedure.The suspect device was returned to the manufacturer for further evaluation.
 
Event Description
It was reported that the roller pump had hesitation when starting the revolutions per minute (rpms) and also had a grinding noise.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician observed the roller pump to operate as intended throughout the evaluation.There was not any noted hesitation when starting or running the pump.
 
Manufacturer Narrative
The reported complaint could not be confirmed.During testing of the device at the service center, the service repair technician observed the roller pump to function properly.Preventive maintenance and release testing was performed.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key11134497
MDR Text Key251999551
Report Number1828100-2021-00009
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801041
Device Catalogue Number801041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2021
Was the Report Sent to FDA? No
Date Manufacturer Received03/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-