Catalog Number 000150 |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device is not expected to be returned for evaluation and review.However, the complaint investigation is not complete at this time.A supplemental and final report will be filed following the completion of the complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
|
|
Event Description
|
The customer reported that the device, 000150, was being used during an esophageal dilation on an unknown date when the device was reported to have bent.There was no report of injury, medical intervention, or hospitalization for the patient.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|
|
Manufacturer Narrative
|
The complaint device was not available for evaluation.No visual evidence (photographs) of failure was made available.A failure analysis of the complaint device could not be completed therefore the reported failure cannot be verified; the complaint is unconfirmed.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A five-year review of complaint history revealed there has been a total of 24 complaints, regarding 24 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.0004.Per the instructions for use, the user is advised that since the marked guidewire is a reusable device that is subjected to varied use and cleaning environments, the life span of the product cannot be guaranteed.In particular, less than 1% of the spring tips have been reported to have become dislodged during reuse or cleaning.Dislodgement of the spring tip during use may require endoscopic removal of the spring tip.Failure to remove the tip may lead to the perforation of the esophagus, stomach or bowel and the consequences customarily associated therewith.Carefully inspect the guidewire after each use.Inspect the flexible spring tip and discard the wire if the tip appears to be bent or fatigued.Also inspect the soldered joint and discard the wire if the soldered joints appear discolored, loose or cracked.This issue will continue to be monitored through the complaint system to assure patient safety.
|
|
Search Alerts/Recalls
|
|
|