On (b)(6) 2020, a 10mm amplatzer septal occluder (aso) was selected for implant.However, the aso deformed into a cobra head shape when deployed in the patient's defect, although several attempts were made.The ado was removed from the patient, it was then the physician tried to reshape the aso, but the axis on the la disk side appeared shifted from the ra disk side.The physician decided it was impossible to restore the shape of the aso, and the aso was replaced with another same-sized aso.The implant procedure was completed safely with no consequence to the patient.
|
The reported event of an amplatzer septal occluder deploying with a cobra deformation could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.Please note, per the amplatzer septal occluder instructions for use, artmt100116885 revision a "caution: do not release the device from the delivery cable if the device does not conform to its original configuration or if device position is unstable or interferes with any adjacent cardiac structure (such as svc, pv, mv, cs, ao).Advance the delivery sheath to recapture the device and redeploy.If the device position is still unsatisfactory, recapture the device and replace it with a new device or abort the procedure.".
|