• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-010
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/21/2020
Event Type  malfunction  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
On (b)(6) 2020, a 10mm amplatzer septal occluder (aso) was selected for implant.However, the aso deformed into a cobra head shape when deployed in the patient's defect, although several attempts were made.The ado was removed from the patient, it was then the physician tried to reshape the aso, but the axis on the la disk side appeared shifted from the ra disk side.The physician decided it was impossible to restore the shape of the aso, and the aso was replaced with another same-sized aso.The implant procedure was completed safely with no consequence to the patient.
 
Manufacturer Narrative
The reported event of an amplatzer septal occluder deploying with a cobra deformation could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.Please note, per the amplatzer septal occluder instructions for use, artmt100116885 revision a "caution: do not release the device from the delivery cable if the device does not conform to its original configuration or if device position is unstable or interferes with any adjacent cardiac structure (such as svc, pv, mv, cs, ao).Advance the delivery sheath to recapture the device and redeploy.If the device position is still unsatisfactory, recapture the device and replace it with a new device or abort the procedure.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key11140871
MDR Text Key225816939
Report Number2135147-2021-00002
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010076
UDI-Public00811806010076
Combination Product (y/n)N
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Replace
Type of Report Initial,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number9-ASD-010
Device Catalogue Number9-ASD-010
Device Lot Number7048370
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-