It was reported that the device morphology diagnostic data was incorrect and that the patient received high voltage therapy inappropriately.Technical support was contacted and reviewed the records from the event.The therapy was triggered by the patient's condition of low far field signal.When the signals are too small, a template cannot be obtained to contribute to the morphology discrimination.The recommended programming was performed to resolve the event.The patient was stable and will continue to be monitored.
|