Model Number 9-ASD-026 |
Device Problems
Material Deformation (2976); Activation Failure (3270)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/12/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Further information regarding this event has been requested.The investigation results will be provided in a subsequent submission.
|
|
Event Description
|
On (b)(6) 2020 during an asd closure procedure, the physician tried to release the device when it got stuck.When the device finally deployed it took on a cobra shape.The patient remained stable.
|
|
Manufacturer Narrative
|
The reported event of the device getting stuck and the device deploying with a cobra deformation could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Manufacturer Narrative
|
The reported event of the device deforming in a cobra shape and difficulty deploying the device could not be confirmed.The investigation confirmed the device met visual, dimensional and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
|
|