• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. XTEND; XTEND ANTERIOR CERVICAL PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. XTEND; XTEND ANTERIOR CERVICAL PLATE Back to Search Results
Device Problem Device Slipped (1584)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
The device was not available for return as it remains in the patient.Imaging provided shows the inferior screw in a 3-level construct has backed out anteriorly past the screw blocking mechanism.The patient is reported to be asymptomatic and there are no plans for a revision surgery.The exact cause of the reported issue could not be determined.
 
Event Description
It was reported that an xtend screw backed out of the plate post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XTEND
Type of Device
XTEND ANTERIOR CERVICAL PLATE
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon 19403
6109301800
MDR Report Key11143142
MDR Text Key227526246
Report Number3004142400-2021-00007
Device Sequence Number1
Product Code KWQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received12/14/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-