Model Number 816300 |
Device Problem
Output Problem (3005)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/17/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the manufacturer's technical support specialist, at first, the system booted up normally.Once it was rebooted and every time thereafter, the error message was displayed.The field service representative (fsr) verified the disk boot failure.He replaced the coin cell battery, which resolved the issue.The hard drive was due to be replaced in six months.The fsr spoke with the facility's biomedical engineer and it was decided to replace the hard drive as a courtesy and a precaution.All release testing was performed.The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) displayed a 'disk boot failure.Insert system disk and press enter' message.As a result, an alternate device was employed.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The product surveillance technician (pst) observed the coin battery to be received with 0.375 volts direct current (vdc).The battery did not have enough voltage to supply power to the basic input output system.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
Additional information received: the surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|