The initial reporter stated they received discrepant results for one patient sample tested with elecsys ferritin on a cobas e 411 immunoassay analyzer.No incorrect results were reported outside of the laboratory.The sample initially resulted with a ferritin value of < 0.500 ng/ml accompanied by a data flag.The sample was repeated, resulting with a value of < 0.500 ng/ml accompanied by a data flag.The sample was sent to a second laboratory for testing on another e411 analyzer, resulting with a value of 10.46 ng/ml.The ferritin reagent lot number was 49581201, with an expiration date of 31-jan-2022.
|
Two reagent packs of ferritin lot 495812 were calibrated on (b)(6) 2020.The first level calibrator signals recovered much higher than expected.The customer stated they were not aware of when the calibrator used for calibration was first opened.Per product labeling: "the calibrators should only be left on the analyzer during calibration at 20-25 °c.After use, close the bottles as soon as possible and store upright at 2-8 °c.Due to possible evaporation effects, not more than 5 calibration procedures per bottle set should be performed.Calibrator stability after opening at 2-8c: 12 weeks." a new calibration was performed on (b)(6) 2020 was ok and signals were only slightly lower than expected.The investigation determined the issue is consistent with a bad calibration due to inappropriate calibrator handling.The issue was resolved by obtaining a new calibration with fresh calibrator material.
|