• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENUSA DE MEXICO S.A. DE C.V. TACTICATH QUARTZ CONTACT FORCE ABLATION CATHETER, 65MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENUSA DE MEXICO S.A. DE C.V. TACTICATH QUARTZ CONTACT FORCE ABLATION CATHETER, 65MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number PN-004 065
Device Problems Display or Visual Feedback Problem (1184); Failure to Analyze Signal (1539)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
One 65 mm tacticath quartz contact force ablation catheter was received for evaluation.Investigation of the returned device revealed the distal shaft was fractured under the strain relief at the distal end of the plunger causing a shift partially over electrode 1.The cause of the fractured shaft is consistent with damage during use.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.Based on the information provided and the investigation performed, the cause of the observed event was determined to be damage during use.
 
Event Description
This report is to advise of an event observed during analysis confirming a displaced electrode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH QUARTZ CONTACT FORCE ABLATION CATHETER, 65MM
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX  32470
Manufacturer (Section G)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX   32470
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key11159325
MDR Text Key228883108
Report Number9680001-2021-00001
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeHU
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/25/2021
Device Model NumberPN-004 065
Device Lot Number6765158
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/04/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-