MAQUET CARDIOPULMONARY GMBH HEATER COOLER UNIT; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HCU 30
Device Problems Electrical /Electronic Property Problem (1198); Failure to Power Up (1476)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/31/2020
Event Type  malfunction  
Manufacturer Narrative
A follow-up emdr will be submitted when additional information becomes available.
 
Event Description
It was reported that the hcu 30 not turn on during a preparation for the emergency surgery.The start of the surgery has been delayed.Complaint number (b)(4).
 
Event Description
Complaint number (b)(4).
 
Manufacturer Narrative
It was reported that the hcu 30 not turn on during a preparation for the emergency surgery.The start of the surgery has been delayed.A getinge field service technician (fst) was sent for investigation on 2021-03-05.The fst confirmed the failure and replaced the mcp00703837 # control board , type 2 pc1891 (material#701034530), the current probe (material#701034527) and the mcp00703831 # power supply board, type 2 (material#701034529 ).The device passed all function tests and has been put back in use.A similar failure was investigated at getinge life cycle engineering on 2017-10-11 as part of a previous complaint.The most probable root cause is a high contact resistance which was probably caused by corrosion or a bad crimp connection.The device was manufactured on 2010-10-28.The review of the non-conformities has been performed on 2021-02-22 for the period of 2010-10-28 to 2020-12-31.It does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences can be excluded.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER COOLER UNIT
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key11160578
MDR Text Key234339702
Report Number8010762-2021-00008
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K031544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Type of Report Initial,Followup
Report Date 05/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHCU 30
Device Catalogue Number70103.4642
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-