Model Number 74120148
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Sleep Dysfunction (2517); Metal Related Pathology (4530)
Event Date 09/02/2020
Event Type  Injury  
Manufacturer Narrative
Complaint reference: (b)(4).
 
Event Description
It was reported that a revision surgery was performed on the patient right hip on (b)(6) 2020.The revision surgery was performed due to difficulty walking and performing other physical activities; loss of range of motion and stiffness, exhaustion from pain, difficulty performing activities of daily life; difficulty sleeping; pain in the right hip; metallosis; adverse reaction to metal debris; and elevated levels of metal ions in blood.Among the intraoperative findings there was a revision acetabular component due to wear of articular bearing surface of internal prosthetic of right hip joint, evidence of trunnionosis and thickened avascular tissue evident.The patient outcome is unknown.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.During the revision, the hemi head and modular sleeve were removed.The bhr cup and stem remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup, hemi head and modular sleeve was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the bhr cup and hemi head.Similar complaints have been identified for the sleeve.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.The clinical information provided, of the trunnionosis and thickened avascular tissue may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
H3, h6: it was reported that right hip revision surgery was performed.During the revision, the hemi head and modular sleeve were removed.The bhr cup and stem remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup, hemi head and modular sleeve was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the bhr cup and hemi head.Similar complaints have been identified for the sleeve.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.The thickened avascular tissue and adverse tissue reaction may be consistent with a reaction to trunnionosis and resulting metal debris.The source and the root cause cannot be determined with the available documentation; however, the patient¿s history of being thrown from a horse cannot be ruled out as a contributing factor to the reported pain and clinical reactions.It cannot be concluded that the reported clinical findings are associated with the implant failure.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, our investigation remains inconclusive and a probable root cause cannot be determined.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 48MM W/ IMPTR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key11167222
MDR Text Key226774740
Report Number3005975929-2021-00001
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010502551
UDI-Public03596010502551
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup,Followup
Report Date 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/29/2013
Device Model Number74120148
Device Catalogue Number74120148
Device Lot Number086192
Was Device Available for Evaluation? No
Date Manufacturer Received05/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-