This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the rpms were at the low end of specification and an aged repair was needed; however, the repair was not completed because the customer declined the repair and the unit was returned to them unrepaired.Review of the previous repair record identified unrelated repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|