Updated fields: d10, g4, g7,h2, h3, h6, h10 unique device identifier (udi) : (b)(4).The certas valve was returned for evaluation: failure analysis - the position of the cam when valve was received was at setting 8.The valve was visually inspected; the silicone housing was cut/torn around the siphon guard and needle holes in the needle chamber were noted.The valve was hydrated.The valve was flushed no occlusion was noted at the ruby ball but leaked from the cut/tear in the silicone housing around the siphon guard.The valve was leak tested and leaked from the cut/tear in the silicone housing around the siphon guard.The valve could not be reflux tested due to the cut/tear in the silicone housing around the siphon guard.The siphon guard was visually inspected marks were noted in the siphon guard.The valve passed the rest for programming, siphon guard and pressure.The root cause for the issue reported by the customer, "csf retention around wound was observed without clinical symptoms¿ was probably due to the cut/tear in the silicone housing.The root cause for the cut/tear in the silicone housing around the siphon guard was probably caused by a sharp or pointed object coming into contact with the silicone housing, as noted in the "ifu" silicone has a low cut/tear resistance.The root cause for the marks in the siphon guard were probably caused by a sharp or pointed object coming into contact with the siphon guard.
|