Model Number 26926 |
Device Problem
Premature Activation (1484)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/08/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that the stent inadvertently deployed.A 5mm x 150mm x 130cm innova was selected for use in a left leg arteriogram with angioplasty and stenting procedure.During preparation, the device was being flushed on the back table when the stent partially deployed.The device never entered the patient.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Event Description
|
It was reported that the stent inadvertently deployed.A 5mm x 150mm x 130cm innova was selected for use in a left leg arteriogram with angioplasty and stenting procedure.During preparation, the device was being flushed on the back table when the stent partially deployed.The device never entered the patient.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Manufacturer Narrative
|
Device evaluation by mfr: returned product consisted of the innova self-expanding stent system.Visual examination revealed a kink to the outer sheath at the nosecone.The stent was deployed and missing.There was blood present inside the sheath.The proximal section of the pull rack was separated and missing.The distal section was still inside the handle.Inspection of the remainder of the device, revealed no other damage or irregularities.Product analysis found damage which could have contributed to the reported event.
|
|
Search Alerts/Recalls
|