APOLLO ENDOSURGERY ORBERA INTRAGASTRIC BALLOON SYSTEM Back to Search Results
Model Number B-4800
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Vomiting (2144)
Event Date 12/24/2020
Event Type  malfunction  
Manufacturer Narrative
The device was returned to the apollo device analysis laboratory on 07/jan/2021.Analysis of the device is ongoing.A review of the device labeling notes the following: the risk documentation for this device establishes the occurrence ranking for the reported event "inflation" as "remote", which is defined by apollo as.02% -.49% of complaints over units sold.A review for the intragastric balloon products for the failure mode "inflation" is performed during quarterly complaint analysis meetings (cam) and has demonstrated that global balloon inflation rate remains at "remote".Therefore, apollo determined that the reported event is occurring within the range of the expected frequency and severity for this reported event, as referenced in the cam meeting slides.The current orbera® intragastric balloon system directions for use (dfu) addresses the known and anticipated potential events of "pain", "vomiting" as follows: "the physiological response of the patient to the presence of the orbera system balloon may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response." "each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms." "complications: possible complications of the use of the orbera system include: gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon.Continuing nausea and vomiting.This could result from direct irritation of the lining of the stomach or as a result of the balloon blocking the outlet of the stomach.It is even theoretically possible that the balloon could prevent vomiting (not nausea or retching) by blocking the inlet to the stomach from the esophagus." orbera® system placement and inflation (step-by-step) slowly fill the balloon with sterile saline, 50 cc at a time.Repeat up to 700 cc (14 strokes).Recommended full volume is up to 700 cc.Minimum fill volume is 400 cc.
 
Event Description
The patient presented with vomiting.The balloon hyperinflated 5 months after placement.
 
Manufacturer Narrative
Supplement 1 medwatch submitted to the fda on 03/22/2021 device evaluation summary: the device was returned to the apollo device analysis laboratory on 7/jan/2021.A deflated balloon with a significant amount of discoloration was returned.There is a large amount of unknown fabric present on the shell.Under microscopic analysis, there is a large amount of fungus present on the shell.Due to the state the balloon was returned in, a functional test could not be conducted.The complaint could not be verified as it is uncertain how the inflation occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBERA INTRAGASTRIC BALLOON SYSTEM
Type of Device
INTRAGASTRIC BALLOON
Manufacturer (Section D)
APOLLO ENDOSURGERY
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX 78746
MDR Report Key11204820
MDR Text Key241612585
Report Number3006722112-2021-00003
Device Sequence Number1
Product Code LTI
UDI-Device Identifier10811955020619
UDI-Public10811955020619
Combination Product (y/n)N
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/06/2022
Device Model NumberB-4800
Device Catalogue NumberB-4800
Device Lot NumberAF03562
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2021
Date Manufacturer Received12/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age29 YR
Patient Weight73
-
-