U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/07/2021
Event Type  Injury  
Event Description
It was reported the patient's ipg no longer accepts a charge, the patient still has therapy.Surgical intervention may be pending to address the issue.
 
Manufacturer Narrative
The reported even for inoperable ipg was not confirmed.At receipt, the ipg successfully communicate with lab utilities.The return ipg accepted charge and was fully recharge at lab charging bank.It also passed all functional testing (bench and autotest).No root cause was identified for the reported issue.
 
Event Description
It was reported surgical intervention was undertaken wherein the ipg was explanted and replaced.Surgical intervention addressed the issue.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
MDR Report Key11204901
MDR Text Key227865956
Report Number1627487-2021-00429
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05414734402651
UDI-Public05414734402651
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/28/2015
Device Model Number3788
Device Catalogue Number3788
Device Lot Number3923834
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2021
Was the Report Sent to FDA? No
Date Manufacturer Received03/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age82 YR
Patient Weight57
-
-