(b)(4).Batch #: u95198.Component code: mechanical (g04).Investigation summary: the analysis results of the el5ml found that the device was received with the jaws closed.The device was disassembled and one jaw, broken at bifurcation, was observed.Further analysis shows evidence of corrosion was found throughout the broken area.In addition, the instrument was found to be empty.The reported complaint was confirmed.The most likely reason for jaw bifurcation breakage is stress corrosion cracking and the most likely root cause is exposure to a solution containing chlorine.In addition, in order to avoid this kind of issues please do not reuse, reprocess, or re-sterilize device.Reuse, reprocessing, or re-sterilization may compromise the structural integrity of the device and/or lead to device failure that in turn may result in patient injury, illness, or death.Please reference the instructions for use for more information.It should be noted that as part of our quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device lot/batch number, and no non-conformances were identified.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
It was reported that during a lap cholecystectomy procedure, on the third clip deployment, the clip would not advance into the jaws and close while attempting to place a clip on the cystic duct (clip failed to engage).They then opened a second el5ml to successfully complete the case.There was no patient consequence.
|