Model Number 8300AB
Device Problem Perivalvular Leak (1457)
Patient Problem Aortic Regurgitation (1716)
Event Date 09/16/2020
Event Type  Injury  
Manufacturer Narrative
Udi (b)(4).The device was not returned to edwards for evaluation.Regurgitation is considered to be a perivalvular leak (pvl) if a turbulent eccentric jet originates between the bioprosthetic sewing ring and the annulus.While the majority of affected patients are asymptomatic, it can lead to significant morbidity including heart failure and hemolytic anemia.Pvl can occur in the mitral and aortic position for similar reasons, including technique and patient related factors.The type and cause of regurgitation varies depending upon multiple factors.Typically, mild regurgitation is not unusual after initial valve replacement, and is usually tolerated by patients.Often moderate to high regurgitation requiring re-operation in the immediate post-operation period is due to procedural related issues and is unrelated to the device.Regurgitation which develops progressively over time can be due to number of issues including patient related factors or structural valve deterioration.A definitive root cause could not be determined.If additional information is received a supplemental mdr will be submitted.Edwards lifesciences will continue to monitor all reported events.No further actions are required at this time.
 
Event Description
It was reported that a patient with a 25mm valve implanted for one year, 11 months underwent balloon valvuloplasty due to perivalvular leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE AORTIC VALVE
Type of Device
TISSUE HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
amritha srinivasan
1 edwards way
irvine, CA 92614
9492504062
MDR Report Key11206379
MDR Text Key227921004
Report Number2015691-2021-00212
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/28/2021
Device Model Number8300AB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/29/2020
Date Device Manufactured08/29/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age70 YR
Patient Weight96
-
-