E1: initial reporter facility name - (b)(6) hospital.D4: lot number - updated to 25644208.Device evaluated by mfr.: returned product consisted of an angiojet solent omni catheter.The pump assembly, effluent/supply line, shaft, tip and spike line were visually examined for damage or any irregularities.The shaft showed a kink located 92.5cm from the tip.The pump and catheter were primed, and the device was run for a period of 120 seconds in the thrombectomy mode.The devices pressure was within the normal range.Power pulse mode was activated, and the device functioned as designed.A.035 test guidewire was inserted into the catheter from the tip and transcended through the device with a slight restriction at the kinked area of the catheter shaft; however, the guidewire transcended through the entire catheter shaft.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|