BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G247
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2020
Event Type  malfunction  
Manufacturer Narrative
This product has not been returned at this time.If it is returned, analysis will be performed, and this report will be updated.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was interrogated prior to implant and was found to be operating in safety mode.The device was therefore not implanted.There was no patient involvement.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was interrogated prior to implant and was found to be operating in safety mode.The device was therefore not implanted.There was no patient involvement.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode and that both brady and tachy therapy remained available.Review of device memory identified an error.The error resulted in software resets performed in an attempt to correct an identified memory inconsistency.The corrupted memory was corrected in the laboratory and the device reverted to normal operation.The cause of the memory inconsistency was not able to be determined through laboratory testing but was likely the result of exposure to radiation, either therapeutic or environmental.This product has not been returned at this time.If it is returned, analysis will be performed, and this report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11212561
MDR Text Key228161546
Report Number2124215-2021-01102
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/27/2022
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number244328
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2021
Date Manufacturer Received12/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/07/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-