MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; IMPLANTABLE LEAD Back to Search Results
Model Number 4542
Device Problems Failure to Capture (1081); High Capture Threshold (3266)
Patient Problems Muscle Stimulation (1412); Undesired Nerve Stimulation (1980); Therapeutic Effects, Unexpected (2099); Inadequate Pain Relief (2388); No Code Available (3191)
Event Date 12/03/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.(b)(4) captures the reportable event of surgery.
 
Event Description
It was reported that this left ventricular (lv) lead was surgically abandoned and replaced during a follow-up, due to high pacing thresholds leading to ineffective stimulation in any pacing configuration and loss of capture.High pacing thresholds were suspected to be caused by patient condition as lv lead parameters were all stable.Also, the patient experienced inadequate stimulation as the phrenic nerve was stimulated generating undesired sensations.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11213681
MDR Text Key228162719
Report Number2124215-2020-28650
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526410901
UDI-Public00802526410901
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/16/2014
Device Model Number4542
Device Catalogue Number4542
Device Lot Number155833
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/16/2012
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age58 YR
-
-