Model Number P3200K000500
Device Problem No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Date 01/19/2021
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the external alarm needed to be replaced.Per the hillrom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Make sure the brake is not set, and then plug the bed into an applicable power source.A steady alarm comes on.Set the brake.The alarm stops.Repair or replace as necessary.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the external alarm to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the bed had no audible brake not set alarm.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
harish vishwanathan
1069 state route 46 east
batesville, IN 47006
3127289851
MDR Report Key11217503
MDR Text Key229356589
Report Number1824206-2021-00025
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP3200K000500
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/19/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-