Model Number G447
Device Problem Defective Device (2588)
Patient Problems Death (1802); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/10/2020
Event Type  Death  
Event Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator expired.A patient advocate noted that the device, in their words, had failed.No further information was provided.This device has not been returned for analysis.
 
Manufacturer Narrative
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Event Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator expired.A patient advocate noted that the device, in their words, had failed.No further information was provided.This device was later returned for analysis with three unidentified leads attached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11219895
MDR Text Key257501728
Report Number2124215-2021-01603
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589539
UDI-Public00802526589539
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/21/2021
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number229733
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2021
Date Manufacturer Received02/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age50 YR
Patient SexMale
-
-