MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problem Poor Quality Image (1408)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/21/2020
Event Type  malfunction  
Manufacturer Narrative
The referenced scope was returned to the olympus service center.The evaluation confirmed the ultrasound image has multiple broken elements.The forceps cover adhesive, objective lens adhesive and light guide lens adhesive at the distal end are peeling off.The electronic connector was found leaking.There are minor buckles on the elevator plug is loose, the scope connector is loose, el connector missing lockpin and the control knob has play/loose.The scope was last repaired on march 26, 2019 due to a leak in the channel.The scope was repaired to standard specifications and returned to the customer.The investigation is ongoing; therefore, the root cause of the reported malfunction cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The user facility reported that during an unspecified event, the evis exera ii ultrasound gastro-videoscope had a "bad picture".No patient injury or harm was reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the forceps cover glue peeling was likely caused by an external force was applied to the distal end leading to peeling of the glue.Additionally, the amount of use and age of the device (over 5 years) is likely to lead to deterioration.This issue is addressed in the instructions for use (ifu) and can be detected: "inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.".Per the legal manufacturer, the other device defects included in the initial mdr have no potential to cause or contribute to death or serious injury if the malfunctions were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11228562
MDR Text Key241236282
Report Number8010047-2021-01880
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170355479
UDI-Public04953170355479
Combination Product (y/n)N
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 03/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/25/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-