Catalog Number 261221 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported that the perforator did not stop and would have plunged during a craniotomy if the surgeon did not stop it.The drill used with the perforator was anspach.
|
|
Manufacturer Narrative
|
The perforator was returned for evaluation.Device history record (dhr): there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis: the perforator unit was inspected using the unaided eye.Unit was soiled with organic matter.No additional anomalies.Ifu testing procedure was performed with no observed anomalies.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release.The unit was found to perform as intended and fulfilled the acceptance criteria.The complaint could not be verified through failure analysis.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.
|
|
Event Description
|
N/a.
|
|