Model Number MSM20 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Attempts are being made to retrieve the device.To date, no additional information has been received and no device has been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported that during a right trans carotid artery procedure, after the first firing, the handles wouldn't open beyond a certain point.A competitor device was used to complete the procedure.There were no adverse consequences for the patient.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 2/23/2021.D4: batch # u94h37.H6: component codes: others (g07).Investigation summary: the product was returned for evaluation.Visual inspection and functional testing was conducted on the returned device.Visual analysis of the returned sample revealed that the msm20 device was returned with no apparent damage.The tyvek was also returned along with the device.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and it fed and formed the remaining eighteen clips as intended.The event described could not be confirmed as the device performed without any difficulties noted and no product defect was identified.Although no conclusion could be reached regarding the cause of the reported event, the instructions for use do contain the following: "caution: ensure that a clip is in the jaws.Position the jaws so that the clip completely surrounds the vessel to be ligated.Fully squeeze the handles together until they stop.As the handles are squeezed, the clip closes and occludes the vessel or tubular structure.Fully release the handles to return the instrument to its original open position.The next clip is automatically advanced.Inspect the jaws after each clip application to ensure the presence of a new clip before applying the next clip.Repeat steps as necessary to continue ligation of tissue.For optimal performance, periodically clean the instrument during the procedure by submerging the jaw and distal end of the applier shaft in saline." as part of our quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
|
Search Alerts/Recalls
|
|
|