Model Number WA47787A |
Device Problems
Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/15/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
|
|
Event Description
|
Olympus was informed that during a therapeutic myomectomy procedure two hf-electrodes of the same lot number broke off and fragments fell inside the patient¿s body cavity.However, no fragments remained inside the patient since they were reportedly retrieved.No further information was provided but there was no report about an adverse event or patient injury.
|
|
Manufacturer Narrative
|
The suspect medical device was not returned to the manufacturer for evaluation/investigation since it was reportedly discarded by the user facility.Therefore, the cause of the reported event could not be conclusively determined.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.The case will be closed on olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
|
|