• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC Back to Search Results
Model Number SPL-T
Device Problems Break (1069); Defective Device (2588)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Device inspection found a broken suction port.Device is non repairable.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
It was reported that during a demo asset return inspection the device was found with broken suction port.There was no patient involvement on this event reported.No user injury reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer¿s final investigation.The device has been previously serviced by olympus, therefore a service history review will replace dhr review.This device has undergone multiple inspections as being asset returns.This device was previously inspected by olympus in january 2020 (ds# 20006886).At that time, minor scratches on the housing were observed.The unit passed functional test, output test, and electrical safety test.Based on the device inspection result, damage to the device was most likely due to user's mishandling.The ifu advises "this product is a precision device; handle it with care.Avoid rough or violent handling, which may cause equipment damage.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY TRANSDUCER.
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough PA 01772
MDR Report Key11251795
MDR Text Key230258192
Report Number3011050570-2021-00035
Device Sequence Number1
Product Code FEO
UDI-Device Identifier00821925043831
UDI-Public00821925043831
Combination Product (y/n)N
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSPL-T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-