This event has been recorded by zimmer biomet under (b)(4).(b)(6).Review of the most recent repair record determined the depth bar had side play, head was worn with visible marks and nicks, and motor, bearings, o-ring, seal, reciprocating arm, head, throttle gasket, and width plate screws were replace and resolved the reported issue.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|