Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).(b)(6).Review of the most recent repair record determined the depth bar had side play, head was worn with visible marks and nicks, and motor, bearings, o-ring, seal, reciprocating arm, head, throttle gasket, and width plate screws were replace and resolved the reported issue.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It was reported that the device was not working properly.Later, during the evaluation of this device, it was determined that the motor speed was running below specification.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11262756
MDR Text Key229838987
Report Number0001526350-2021-00117
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number61153333
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/06/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-