Catalog Number A1059 |
Device Problem
Material Too Rigid or Stiff (1544)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/30/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported that the screw was stiff when tightened on the mayfield skull clamp (a1059).The patient was prepped for a craniotomy procedure.There was no reported delay in surgery and no patient injury.
|
|
Manufacturer Narrative
|
Mayfield modified skull clamp (a1059) was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
|
Manufacturer Narrative
|
The mayfield modified skull clamp (a1059) was returned for evaluation: device history record: the dhr shows no abnormalities related to the reported failure.Failure analysis - unit was received with the lock having both rotational and lateral movement and a residue buildup was present.The plunger assembly requires cleaning.Upon disassembly, repair noted the index knob and lock will need new components added to replace worn internal parts.Unit needs to be machined to have heli-coils added to the large starburst threads.The set screw in the swivel base was tight.General maintenance and cleaning, along with addition of heli-coils to the large starburst threads is also required.Root cause - the reported complaint was confirmed via inspection of the unit.The ratchet extension movement was being impeded by the plunger assembly which needed cleaning.Unit required replacement of worn parts, general maintenance and cleaning.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|
Search Alerts/Recalls
|