• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problem Material Too Rigid or Stiff (1544)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/30/2020
Event Type  malfunction  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the screw was stiff when tightened on the mayfield skull clamp (a1059).The patient was prepped for a craniotomy procedure.There was no reported delay in surgery and no patient injury.
 
Manufacturer Narrative
Mayfield modified skull clamp (a1059) was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Manufacturer Narrative
The mayfield modified skull clamp (a1059) was returned for evaluation: device history record: the dhr shows no abnormalities related to the reported failure.Failure analysis - unit was received with the lock having both rotational and lateral movement and a residue buildup was present.The plunger assembly requires cleaning.Upon disassembly, repair noted the index knob and lock will need new components added to replace worn internal parts.Unit needs to be machined to have heli-coils added to the large starburst threads.The set screw in the swivel base was tight.General maintenance and cleaning, along with addition of heli-coils to the large starburst threads is also required.Root cause - the reported complaint was confirmed via inspection of the unit.The ratchet extension movement was being impeded by the plunger assembly which needed cleaning.Unit required replacement of worn parts, general maintenance and cleaning.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key11263982
MDR Text Key233485052
Report Number3004608878-2021-00044
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 05/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Date Manufacturer Received05/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-